Reliability of functional electromechanical dynamometry
What ICC, standard error and minimal detectable change mean, what the published studies show and how to use those numbers to decide whether someone has improved.
Content by MyoQuality · Updated: 2026-11-21
Reliable is not the same as valid
A device is reliable if measuring the same thing twice under the same conditions gives the same result. It is valid if it measures what it claims to, compared with a reference. A device can be highly reliable and carry a constant bias against another: its data are fine for following a person, but not for mixing with the other device’s data.
Relative reliability is expressed with the intraclass correlation coefficient (ICC): above 0.90 is considered excellent. Absolute reliability uses the standard error of measurement (SEM) in real units or the coefficient of variation (CV) as a percentage.
The number that matters: minimal detectable change
To decide whether someone has improved you need the minimal detectable change (MDC). At 95 % confidence it is SEM × 1.96 × √2. A change smaller than the MDC may be measurement noise.
Example: if the SEM of mean force in a test is 20 N, the MDC95 is about 55 N. A 40 N gain is not enough to say the person is stronger. An 80 N gain is.
That is why it pays to use the test’s most stable variable, which is not always the most eye-catching one, and to familiarise the person before the first baseline.
What the studies show
Squat: very high reliability in strength and endurance protocols, with range of motion and maximal dynamic force the most stable variables over time.
Velocity: near-perfect correlation with linear velocity transducers and very low random error, with a small systematic bias. Valid for velocity-based training.
Isometric mid-thigh pull: reliable for peak force, not suitable for rate of force development, and not interchangeable with force plates.
Sit-to-stand, in healthy young adults: near-perfect reliability for force and velocity with incremental loads of 10 and 15 kg.
Shoulder and trunk: a highly reliable standing isokinetic shoulder rotator test, with 0.6 m/s the recommended velocity; trunk flexors with CVs generally below 10 %; and a multicentre trunk rotator study in which mean force was more consistent than peak force.
Eccentric hamstrings in soccer players: ICCs from 0.87 to 0.91. Isometric and isometric-vibratory protocols in the bench press, row and knee extension: high reliability in both modes.
Rules for your own protocol
Always use the same device, the same position, the same velocity and the same cues. Record anchor heights and range of motion.
Run at least one familiarisation session before the baseline.
Prefer mean force when studies show it is more stable than peak, and the variable with the lowest CV in your population.
Compute your test’s MDC from your own test-retest data when you can. Otherwise, use one from a study with a population similar to yours.